Reliability of normalisation methods for EMG analysis of neck muscles

Work. 2006;26(2):123-30.

Abstract

Acceptable reliability of normalisation contractions in electromyography (EMG) is paramount for testing conducted over a number of days or if normal laboratory strength testing equipment is unavailable. This study examined the reliability of maximal voluntary isometric contractions (MVIC) and sub-maximal (60%) isometric contractions for use in neck muscle EMG studies. Surface EMG was recorded bilaterally from eight sites around the neck at C4/5 level from five healthy male subjects. Subjects performed MVIC and sub-maximal normalisation contractions using an isokinetic dynamometer (ID) and a portable cable dynamometer with attached strain gauge (PCD) in addition to a MVIC against a manual resistance (MR). Subjects were tested in flexion, extension, left and right lateral bending and were retested by the same tester within a two-week period. Intra class correlation co-efficients (ICC) were calculated for each testing method and contraction direction and a mean ICC was calculated across all contraction directions. All normalisation methods produced excellent within-day reliability (mean ICC >0.80) but only the MVICs using the ID and PCD had acceptable reliability when assessed between-days. This study confirmed the validity of using MVICs elicited using the ID and PCD as reliable reference contractions for the normalisation of neck EMG.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adult
  • Electromyography / standards
  • Humans
  • Isometric Contraction / physiology
  • Male
  • Neck Muscles / physiology*
  • Occupational Health
  • Reference Values
  • Western Australia